	 Maximum Loading Capacity On Micrometer Heads

The maximum loading capacity of a micrometer head depends mainly on the method of mounting and whether the loading is static or dynamic (used as a stop, for
example). Therefore the maximum loading capacity of each model cannot be definitively specified. The loading limits recommended by Mitutoyo (at less than 100,000
revolutions if used for measuring within the guaranteed accuracy range) and the results of static load tests using a small micrometer head are given below.

1. Recommended maximum loading limit

B Maximum loading limit
                         Standard type  (spindle pitch: 0.5 mm)                                   Up to approx. 4 kgf *

                                        Spindle pitch: 0.1 mm/0.25 mm                             Up to approx. 2 kgf

                                        Spindle pitch: 0.5 mm                                     Up to approx. 4 kgf

                         High-functionality type Spindle pitch: 1.0 mm                            Up to approx. 6 kgf

                                        Non-rotating spindle                                      Up to approx. 2 kgf

                                        MHF micro-fine feed type (with a differential mechanism)  Up to approx. 2 kgf

                * Up to approx. 2 kgf only for MHT

2. Static load test for micrometer heads (using MHS for this test)

(1) Locknut              (2) Split-body clamp                           (3) Setscrew clamp

             P                                                                                P

                Locknut                                                                           Setscrew  Test method
                                                                                                            Micrometer heads were set up as shown and
                                        P                                                                   the force at which the head was damaged or
                                                     Split-body clamp                                       pushed out of the fixture when a static load
                                                                                                            was applied, in direction P, was measured.
                                                                                                            (In the tests no account was taken of the
                                                                                                            guaranteed accuracy range.)

                                                                        Mounting method                     Damaging / dislodging load*

                                                                        (1) Locknut              Damage to the main unit will occur at 8.63 to 9.8kN (880 to 1000 kgf).

                                                                        (2) Split-body clamp The main unit will be pushed out of the fixture at 0.69 to 0.98kN (70 to 100 kgf).

                                                                        (3) Setscrew clamp Damage to the setscrew will occur at 0.69 to 1.08kN (70 to 110 kgf).

                                                                        * These load values should only be used as an approximate guide.