для размещения заказа - mitutoyo@mitutoyo-tools.ru
EV Counter Functions
SERIES 542 – For Multi-Gauge System GO/±NG judgement, GO/±NG signal output, MAX/MIN/TIR
(runout) measurement, counting direction switching, mm/inch
• Up to six gauges can be connected to one unit. switching, calculation of sum, average, maximum, minimum
• Up to 10 EV counters can be connected to one personal computer using the RS Link function to and maximum difference between specified axes, and output-
facilitate the configuration of a multi-point measurement system comprising a maximum of 60 gauges. mode selection
• A range of output modes are available: I/O output for tolerance judgement and segment output, BCD
data output and RS-232C. Output Normal operation output and GO/±NG signal
• Peak-hold measurements are possible for maximum value, minimum value, runout (TIR), etc. (three steps), measurement data (BCD code),
• Able to display the sum, average, maximum, minimum, maximum difference, etc., between gauges I/O: or 21-stage segment output (selectable)
connected to the same unit. Various measurement data
RS-232C:
External control
I/O: Axis designation, preset, data hold, and error
clear
RS-232C: Displayed value output command, MAX/MIN/
TIR switching and peak value clear, zero set,
preset value input, tolerance value input,
error clear, and command to output
calculated value between specified axes
542-063 542-064E RS link
SPECIFICATIONS Up to 10 EV counters may be connected via a single RS-232C
port (daisy chained). EV and EF counters can be mixed (in
Code No. 542-063 542-064E which case a total of six counters can be connected)
Applicable input Differential square wave Digimatic code (SPC) Error display/output
Applicable gauge LGB (ex. 0.0001 mm resolution), LGF 6 LGS Power-supply voltage error, overspeed error, overflow error,
Number of gauge inputs gauge error, communication error, and tolerance setting error
0.0001**, 0.0005, 0.001, 0.005, 0.01 mm 0.001*, 0.01 mm (.00005"*, .0005")
G Resolution (.000005", .00005", .0005") 927.00 Maximum input frequency
977.00
Price 1.25MHz (differential square wave): max counting speed:
5MHz
* Will not be indicated when combined with an LG gauge.
** D-EV is required when the resolution is 0.0001 mm. Power supply
DIMENSIONS Terminal block (M3 screws), 12 - 24V DC, 700mA (max.)
Unit: mm
4 64 4 6.5
6 72 Interconnecting
EV counters
4.5
4 136 4 5 161 5 Affixing an
144 16 EV counter
139 16
System Diagram
Display unit
EH Counter EV Counter
Liear gauge Programmable
controller
(Tolerance judgement output,
BCD output, segment output,
RS-232C output)
PC system
(RS-232C output)
G-9 Mitutoyo operates a policy of continuous improvement that aims to provide the customer with the benefit of the latest technological advances.
Therefore the company reserves the right to change any or all aspects of any product specification without notice.
mitutoyo@mitutoyo-tools.ru www.mitutoyo-tools.ru