Gauge Blocks
SERIES 516 – Maintenance Kit for Gauge Blocks
• Maintenance kit for gauge blocks includes all the necessary maintenance tools for removing burrs and
contamination before use, and applying anti-corrosion treatment after use, etc.
E Recommendation for Regular
Calibration
516-650E
Gauge blocks are often used to define a company’s standard
of length for manufacturing and as such must be reliable. This
means that they need regular calibration to verify accuracy.
(The problem of damage or corrosion should be addressed
during use and blocks seriously affected must be discarded
immediately.) The frequency of calibration depends on the
tolerance requirements of the work, the amount of use and
conditions under which the gauge blocks are used. The most
economical cycle for any particular set of gauge blocks is best
determined by studying the calibration history. The list below
indicates timings for a typical initial calibration cycle for the
various grades of block.
Application Cycle Grade (reference)
Reference 1-2 K
Standard
Inspection 2 K or 0
Shop floor 2 0 or 1
0.5 - 1 1 or 2
As an accredited calibration laboratory, Mitutoyo offers a
traceable calibration service for customers’ gauge blocks. Our
regular calibration service features:
SPECIFICATIONS • Gauge blocks manufactured by any maker can be
calibrated.
Code No. Tools and accessories included Price • Cleansing and removal of burrs.
516-650E Ceraston (601645): For removing burrs on the measuring surfaces (100 x 25 x 12 mm). • Central dimension and dimensional deviations of each block
Optical flat (158-117): For checking whether burrs exist on the measuring surfaces
(45 x 12 mm). are measured.
Tweezers (600004): For handling thin blocks.
Blower brush (600005): For blowing dust off the measuring surfaces. • Calibration results are provided for immediate use and for
Cleaning paper (600006): For wiping off rust preventive oil and contamination (500 pcs.).
Artificial leather mat (600007): For handling blocks on a benchtop to prevent contamination building a calibration history of each block.
and damage
Reagent bottle (600008): Bottle of cleansing solution (100 ml). 462.00
Gloves (600009): For handling large blocks to prevent expansion due to heat transfer from
operator.