для размещения заказа - mitutoyo@mitutoyo-tools.ru
Digimatic Indicators Technical Data
SERIES 543 – ABSOLUTE Digimatic Signal Output Indicator ID-C Display: 6-digit LCD and sign
• With the MAX/MIN value holding function, this Scale type: ABSOLUTE electrostatic linear
indicator can output the signal of a GO/±NG
judgement result against the peak value set. encoder
• Electronic processing provides high reliability
and accuracy. The seven-core flying lead cable Stem dia: 8 mm
can be used by the customer to connect to
external equipment as required. (See the I/O Standard contact point: 901312
Specifications table for signal output and power
supply details, and the Output/Input circuit Dust/water protection
diagrams for typical interface details.)
• The ABSOLUTE sensor restores the last level: IP54
origin position automatically when the indicator
is turned on. It also realizes high reliability by Functions
eliminating over-speed errors.
• Provided with a 4 m cable. Signal output (–NG/OK/+NG, NPN open collector), Remote
• External power 12–24VDC required. control (hold-reset, preset-recall, zero-set), Preset, Zeroset,
GO/±NG judgement, Max/Min/Runout value holding,
543-281B Counting direction switching, Power ON/OFF, inch/mm
reading
F Alarm: Counting value composition error, Overflow
error, Tolerance limit setting error
Optional Accessories
902011: Plunger lifting lever
2.80
540774: Plunger lifting cable
20.50
Contact points for Mitutoyo's dial indicators (refer to pages
F-37 to F-40 for details)
Interchangeable backs (refer to page F-41 for details)
Measuring stands (refer to pages F-70 to F-73 for details)
Output pattern
Wire – NG OK + NG Composition error
Orange (– NG) Low High High High
SPECIFICATIONS Green (OK) High Low High High
Brown (+ NG) High High Low High
Inch/Metric LED Red Green Red Red (blinking)
Code No. Range Resolution Accuracy* Measuring force Price LCD < O > “x.xxE” indication
543-281B .5"/12.7 mm 420.00
.00005"/0.001 mm .00012"/0.003 mm 2.0 N or less I/O Specifications
* Quantizing error of ±1 count is excluded. Wire Signal I/O Description
DIMENSIONS Black – V (GND) — Connected to minus (-)
terminal
20 28.5
Red + V I Power supply (12 - 24VDC)
Unit: mm Orange – NG O Tolerance judgement
20 Green OK result output:
Brown + NG Only the terminal
O corresponding to a
10.9 Cable: 4 m judgement result is set to
O the low level.
Yellow PRESET_RECALL I External input terminal:
ZERO If the relevant terminal
54 Blue HOLD_RESET I is set to the low level, its
6.5 signal becomes true.
60.55 Shield FG — Connected to GND
13.3 67.4 Note: Measurement data cannot be output.
43.5 50.1 Output OUT
37.2 13.5 circuit
7.3
1000pF Load
8 -00.009 SG
7.6 Output voltage: Max. 24V
Output current: Max. 30mA
Output saturation voltage: Max. 0.3V
Input DC+12-24V Use a non-
circuit voltage output
such as an open
24kW collector output
1000pF OUT or relay output.
Input current: SG
Max. 10mA
F-13 Mitutoyo operates a policy of continuous improvement tIhDa-tCa1i1m2sJ to provide the customer with the benefit of the latest technological advances.
Therefore the company reserves the right to change any or all aspects of any product specification without notice.
mitutoyo@mitutoyo-tools.ru www.mitutoyo-tools.ru