RS Link Function Multi-point measurement can be performed by connecting multiple EH or EV counters with RS Link cables.
RS Link for EH Counter
It is possible to connect a maximum of 10 counter units and handle up to 20 channels of multi-point measurement at a time.
For this connection use a dedicated RS Link cable (02ADD950, 0.5 m; 936937, 1 m; 965014, 2 m).
(The total length of RS Link cables permitted for the entire system is up to 10m.)
First counter Last counter
IN OUT IN OUT IN OUT
RS-232C connector RS-232C connector RS-232C connector
RS-232C cable *
Personal computer
*Only the Mitutoyo software
“SENSORPAK” can be used
when a USB cable is connected.
Gauge number 01 02 03 04……
RS Link for EV Counter G
It is possible to connect a maximum of 10* counter units and handle up to 60 channels of multi-point measurement at a time.
For this connection use a dedicated RS Link cable (02ADD950, 0.5 m; 936937, 1 m; 965014, 2 m).
(The total length of RS Link cables permitted for the entire system is up to 10 m.)
* The maximum number of counter units that can be connected is limited to 6 (six) if an EH counter is included in the chain.
Unit 1 Unit 2
IN OUT
IN OUT IN OUT RS-232C connector IN OUT
RS-232C connector External display unit 1 External display unit 2
……
RS-232C cable
Personal computer
……
Gauge number 01 06 07 12……